Chemical Oxygen Demand (COD) measures how much oxygen it would take to chemically oxidize the organic matter in a sample. It is reported in mg/L O2. Treatment plants, industrial dischargers and environmental labs use COD to track organic load, check how well treatment is working and meet discharge permits.
COD is popular partly because it is fast. A BOD test takes five days. A COD result is ready in a few hours.
How the vial (closed reflux) method works
Add the sample to a pre-filled COD vial that contains a strong oxidant (potassium dichromate) in sulfuric acid, along with catalysts.
Digest the sealed vial in a heating block, typically at 150 °C for two hours.
Cool the vials to room temperature.
Read the result in a photometer, colorimeter or spectrophotometer that is programmed for the COD range you use.
Pre-filled vials take away the hazards and the guesswork of mixing reagents yourself. That is why most labs use them today.
The equipment you need
1. A COD reactor (digestion block)
Hach COD Digital Reactor Block
The Hach COD Digital Reactor Block keeps vials at the correct digestion temperature and includes timer functions. Pick a block size that matches your daily sample volume. For storing and cooling vials, a 40-count COD vial rack keeps the bench organized.
2. An instrument to read the vials
Hach DR1900Hach DR3900CHEMetrics COD photometer
Hach spectrophotometers: the portable DR1900 or the benchtop DR3900. Both run COD alongside many other methods. Older Hach colorimeters may need a COD vial adapter.
COD vials come in several ranges, such as low range for treated effluent and high range for raw influent and industrial wastewater. Choose a range that puts your expected results in the middle of the scale. If a sample reads near the top, dilute it and run it again.
4. Standards for QA
Run a standard of known value with every batch. CHEMetrics COD calibration standards (also available as a second range) let you verify both your digestion and your reading. Keep a record of the results. Auditors will ask for it.
Best practices for reliable COD results
Homogenize samples that contain solids before you pipette. A laboratory blender helps with difficult samples.
Pipette accurately. Small errors in sample volume become large errors in the result. Use calibrated pipettes and clean tips.
Watch out for chloride. High chloride can interfere. Check the method's chloride limit and dilute or use a chloride-tolerant method where needed.
Let vials cool completely and wipe them clean before reading.
Run a blank from the same vial lot as your samples.
Safety and disposal
COD vials contain concentrated acid and hexavalent chromium, and many also contain mercury compounds. Wear eye protection such as safety spectacles and gloves. Never open hot vials. Dispose of used vials as hazardous waste under your local regulations.
Equip your COD lab with Ramo Trading
Whether you are setting up a new wastewater lab or replacing a reactor, Ramo Trading supplies Hach and CHEMetrics COD equipment and consumables. Contact us for a complete package quote.
A practical overview of the leading water-testing brands, what each does best, and how to build the right testing setup for drinking water, wastewater, industrial water,...
Ramo Trading has published a series of practical buying and application guides for water and wastewater testing, covering Hach, CHEMetrics, LaMotte, Palintest, IDEXX and more.
Total coliforms and E. coli are the main indicators of microbiological water safety. Compare the two most common test methods, IDEXX Colilert enzyme substrate testing and...
Comments (0)