Turbidity Measurement Guide: Getting Accurate Results with the Hach 2100Q

By Ramo Trading 3 min read

Turbidity is the cloudiness of water caused by suspended particles such as clay, silt, algae, organic matter and microorganisms. It matters for more than appearance. High turbidity can shield pathogens from disinfection, and it often signals a problem with filtration or the source water. That is why turbidity is a closely regulated parameter in drinking water treatment.

NTU vs. FNU: why the units differ

Turbidimeters measure light scattered by particles, usually at 90° to the light beam. The unit depends on the light source and the standard the method follows:

  • NTU (Nephelometric Turbidity Units) is associated with a white-light (tungsten) source, as specified in US EPA Method 180.1, which US drinking-water reporting commonly uses.
  • FNU (Formazin Nephelometric Units) is associated with an infrared (860 nm) LED source, as specified in ISO 7027, which is common in Europe and many international markets.

On the same sample, the two methods can give different numbers, especially if the sample is colored. Choose the method your regulator or customer expects.

Hach 2100Q vs. 2100Qis

Hach 2100Q portable turbidimeter
Hach 2100Q (EPA method)
Hach 2100Qis portable turbidimeter
Hach 2100Qis (ISO 7027)

Hach's portable turbidimeters bring lab-grade turbidity measurement to the field:

  • Hach 2100Q: the EPA-method version with a tungsten light source, typically chosen by US utilities.
  • Hach 2100Qis: the ISO 7027 version with an infrared LED, often preferred internationally and for colored samples.

Both offer guided calibration, data logging and a rugged portable design. For a wider choice of instruments, browse our turbidimeters collection.

Sample cells: where accuracy is won or lost

At low turbidity, the sample cell often causes more error than the instrument does. Follow these rules:

  1. Hold cells only by the top, never by the light-path area.
  2. Clean and dry the outside with a lint-free cloth before every reading.
  3. Apply silicone oil as the manufacturer directs, to mask small scratches.
  4. Index the cell. Mark its orientation and always insert it the same way.
  5. Remove bubbles by letting the sample rest, or by gentle degassing. Never shake the sample.
  6. Replace scratched cells. For colorimetric work, Hach also offers cells such as the 1-inch square 25 mL glass sample cell.

Calibration and verification

  • Calibrate with formazin or approved primary standards at the interval your method requires.
  • Use secondary standards (gel or sealed standards) between calibrations to verify performance quickly.
  • Record every calibration and verification. Inspectors will expect to see it.

Measure samples promptly

Turbidity can change after sampling as particles settle, flocculate or grow. Measure as soon as possible. Before you fill the cell, mix the sample gently by inverting the container rather than shaking it.

Beyond turbidity

Hach DR300 chlorine colorimeter
DR300 chlorine
Hach Pocket Pro+ pH tester
Pocket Pro+ pH

Many operators pair turbidity with chlorine and pH checks on the same rounds. A DR300 chlorine colorimeter and a Pocket Pro+ pH tester complete a compact field kit.