Refractive Index Detector RI 2012 detectors are available for:
RI 2012 M / micro | RI 2012 A / analytical | RI 2012 P/ semiprep | |
Detection Method | Deflection Deflection Deflection | ||
Refractive Index Range | 1.00 to 1.75 1.00 to 1.75 1.00 to 1.75 | ||
Flow Rate | 0,2 - 3,0 ml/min | 0,2 - 3,0 ml/min | 5,0 - 50 ml/min |
Flow Cell Volume | 4ul, 45 degree angle | 9ul, 45 degree angle | 13ul, 5 degree angle |
max. Flow Cell Pressure | 6 kg/cm2 6 kg/cm2 6 kg/cm2 | ||
Dead Volume | Into cell 6ul | Into cell 24ul | Into cell 88 or 314ul |
Linearity Range | 0-500ul RIU | 0-1000 uRIU | 0-20000 uRIU |
Noise Level | 10 x 10-9 RIU | 5 x 10-9 RIU | 10 x 10-8 RIU |
Autozero Range | Full Range | Full Range | Full Range |
Drift with 1ml H2O/min | < 1 mV/hour | < 1 mV/hour | < 1 mV/hour |
Purge Valve | yes | yes | Yes/No (depends on flow) |
Integrator Output | +/- 1 Volt | ||
Recorder Output | +/- 10 mV / 100 mV / 1Volt | ||
Recorder Offset | 0 mV/ 10 mV/ 100 mV | ||
Recorder Range | 8 steps (1:8) - (16:1) | ||
Marker | Yes / No | ||
Digital Interface | Control of: Purge, Autozero, Start, Stop, Heater, Data output rate, Polarity Current output of: Temperature optical bench, Optical balance, sum voltage, difference voltage, Alarm (Signal too high/low, A/D out of range, no temp. sensor) | ||
Digital Output | TTL: Intensity Alarm | ||
Digital Input | TTL: Purge, Autozero, Start/Marker | ||
Temperature Setting | Ambient, 35 degree C to 55 degree C in 1 degree C steps, Thermal Fuse 70 degree C | ||
Time Constant | RAW (0,0 sec), Fast (0,4 sec), Medium (0,8 sec), Slow (1,2 sec) | ||
Power Source | AC 100-120/220-240V, 50/60 Hz, 50VA | ||
Dimensions in cm | 300mm x 440mm x 175mm (W x D x H) | ||
Weight | 9,6 kg |
Cat.-No.: | Description |
RI 2012 A | Differential Refractive Index Detector, Analytical |
RI 2012 M | Differential Refractive Index Detector, Micro |
RI 2012 P | Differential Refractive Index Detector, Preparative |