COD Testing Made Simple: Equipment Guide for Wastewater Labs

By Ramo Trading 3 min read

Chemical Oxygen Demand (COD) measures how much oxygen it would take to chemically oxidize the organic matter in a sample. It is reported in mg/L O2. Treatment plants, industrial dischargers and environmental labs use COD to track organic load, check how well treatment is working and meet discharge permits.

COD is popular partly because it is fast. A BOD test takes five days. A COD result is ready in a few hours.

How the vial (closed reflux) method works

  1. Add the sample to a pre-filled COD vial that contains a strong oxidant (potassium dichromate) in sulfuric acid, along with catalysts.
  2. Digest the sealed vial in a heating block, typically at 150 °C for two hours.
  3. Cool the vials to room temperature.
  4. Read the result in a photometer, colorimeter or spectrophotometer that is programmed for the COD range you use.

Pre-filled vials take away the hazards and the guesswork of mixing reagents yourself. That is why most labs use them today.

The equipment you need

1. A COD reactor (digestion block)

Hach COD digital reactor block
Hach COD Digital Reactor Block

The Hach COD Digital Reactor Block keeps vials at the correct digestion temperature and includes timer functions. Pick a block size that matches your daily sample volume. For storing and cooling vials, a 40-count COD vial rack keeps the bench organized.

2. An instrument to read the vials

Hach DR1900 portable spectrophotometer
Hach DR1900
Hach DR3900 spectrophotometer
Hach DR3900
CHEMetrics COD photometer
CHEMetrics COD photometer

3. Vials in the right range

COD vials come in several ranges, such as low range for treated effluent and high range for raw influent and industrial wastewater. Choose a range that puts your expected results in the middle of the scale. If a sample reads near the top, dilute it and run it again.

4. Standards for QA

Run a standard of known value with every batch. CHEMetrics COD calibration standards (also available as a second range) let you verify both your digestion and your reading. Keep a record of the results. Auditors will ask for it.

Best practices for reliable COD results

  • Homogenize samples that contain solids before you pipette. A laboratory blender helps with difficult samples.
  • Pipette accurately. Small errors in sample volume become large errors in the result. Use calibrated pipettes and clean tips.
  • Watch out for chloride. High chloride can interfere. Check the method's chloride limit and dilute or use a chloride-tolerant method where needed.
  • Let vials cool completely and wipe them clean before reading.
  • Run a blank from the same vial lot as your samples.

Safety and disposal

COD vials contain concentrated acid and hexavalent chromium, and many also contain mercury compounds. Wear eye protection such as safety spectacles and gloves. Never open hot vials. Dispose of used vials as hazardous waste under your local regulations.