Parameter | 6850 |
---|---|
Wavelength range | 190 to 1100nm |
Wavelength resolution | 0.1nm |
Wavelength accuracy | ? 0.3nm (at 0.5 and 1nm bandwidth) |
Wavelength reproducibility | ?0.2nm |
Spectral bandwidth | Variable 0.5, 1, 2, 4, 5nm |
Photometric range | -0.3 to 3.0A, 0 to 200T |
Photometric accuracy | ? 0.002A (0-0.5A), ?0.3T (0-100T) |
Photometric reproducibility | ?0.001 Abs (0 to 0.5 Abs), ?0.002 Abs (0.5 to 1.0 Abs), 0.15T (0-100T) |
Resolution | 0.1T, 0.001A |
Stray light | 0.05T at 220 and 360nm |
Noise | 0.0005A |
Stability | ? 0.001Ah at 500nm after 15 min warm up |
Multi-wavelength | Up to 10 wavelengths, up to 20 wavelengths with PC software |
Calculations | Ratio, difference, formulae with factors |
Spectrum range | Any range between 190 and 1100nm |
Scan speed | 100 to 2000nm min |
Scan interval | 0.1, 0.2, 0.5, 1, 2 or 5nm |
Kinetics | Up to 12 hours with time intervals of 0.1, 0.2, 0.5, 1, 2, 5, 10 or 30 seconds |
Analysis | Slope and formula of line of best fit between any two points |
Quantitation points | Up to 3 wavelengths |
Quantitation Calibration | Blank with up to 10 standards or factor |
Concentration range | 0-99999 |
Calibration | Blank with standards or factor |
DNA | DNA Ratio, concentration, A320 correction |
Light source | Tungsten halogen and Deuterium lamps |
Lamp changeover | 325 to 370nm selec table |
Outputs | USB and parallel |
Operating system: | Windows 2000, XP, Vista, Windows 7 |
Electrical supply | 120VA, 220110V, 5060Hz |
Size (w x d x h) | 600 x 450 x 200mm (23.6 x 17.7 x 7.9in) |
Weight | 22 kg (48.4 lbs) |
Couldn't load pickup availability